Keysight E5071C ENA Network Analyzers 31
General Information
Power Meter Accuracy Test
Power Meter Accuracy Test
This test is intended for power meters used in testing the E5071C. The
“Dynamic Accuracy Test” requires the use of a power meter that has been
calibrated to a higher accuracy than the standard power meter.
Power meters with options G12 and H12 specify an improved instrument
accuracy over a limited power range. (These power meters do not contain
unique hardware.) A power meter may be returned to the factory to have one of
these options added to an existing power meter or to renew the calibration for
one of these options.
This test procedure is an alternative to returning the power meter to the
factory. When a power meter passes this test, it is considered to be calibrated
for the G12 or H12 option even though it has not been returned to the factory.
Power Meters That Can Be Tested Using This Procedure
This procedure assume that the E4419B power meter is being tested.
It is recommended that the revision number for the power meter “Main
Firmware” be Ax.03.00 or higher.