Keysight E5071C ENA Network Analyzers 275
Replacement Procedure
LCD Replacement
Figure 6-41 LCD Removal
Replacement Procedure
Step 1. Replace the LCD by inverse procedure of removal.
Step 2. You have to calibrate the touch screen. Follow the procedure described on
“Calibration of the Touch Screen” on page 208.
Fasten the TORX T15 screws (item 5) using a torque screwdriver (set to
1.02 N-m / 9.0 lb-in).