214 Keysight E5071C ENA Network Analyzers
Replacement Procedure
Hard Disk Drive Replacement Option 019
6-
Hard Disk Drive Replacement Option 019
Tools Required
TORX screwdriver, T10
Torque screwdriver, TORX T10 (set to 0.50 N-m / 4.4 lb-in)
Removal procedure
Refer to Figure 6-10 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 203.
Step 2. Disconnect the two cables (item 1).
Step 3. Remove the four TORX T10 screws (item 2) fastening the hard disk assembly.
Step 4. Remove the four TORX T10 screws (item 3) fastening the hard disk bracket.
Step 5. Remove the hard disk drive (item 4) from the connector board.
Figure 6-10 Removable Hard Disk Drive Removal Option 019
From the volume label CN925, user recovery function is not prepared. It is