36 Keysight E5071C ENA Network Analyzers
General Information
Power Meter Accuracy Test
1-
Test Record for Power Meter Accuracy Test
Test Equipment Used
Test Results
Pass/Fail Calculation
Power Meter Tested
Model Number: Test Date:
Serial Number: Tested by:
Range Calibrator: Model No.: Serial No.:
Digital Voltmeter: Model No.: Serial No.:
Range
Calibrator:
Setting
Channel A Channel B
DVM Reading
(Vdc)
Power Meter
Reading (dBm)
DVM Reading
(Vdc)
Power Meter
Reading (dBm)
1 mW A= 0.000.001 A= 0.00 0.001
300 W B = C = B = C=
100 WD =E= D =E =
300 WR=B/A =
S=10
(C/10)
=
% ERROR=((R-S)/R)x100 =
Limits: 0.13%
Pass __ Fail __
R=B/A =
S=10
(C/10)
=
% ERROR=((R-S)/R)x100 =
Limits: 0.13%
Pass __ Fail __
100 WT=D/A =
U=10
(E/10)
=
% ERROR=((T-U)/T)x100 =
Limits: 0.10%
Pass __ Fail __
T=D/A =
U=10
(E/10)
=
% ERROR=((T-U)/T)x100 =
Limits: 0.10%
Pass __ Fail __