66 Keysight E5071C ENA Network Analyzers
Troubleshooting
How to exit from the E5071C Measurement View
4-
How to exit from the E5071C Measurement View
You need to exit from the E5071C Measurement View to perform some
troubleshooting. The following is the procedure to exit from the E5071C
Measurement View.
Step 1. Connect the mouse and external keyboard to the connectors on the E5071C
rear panel.
Step 2. Turn the instrument on.
Step 3. Press key.
Step 4. Click Service Menu - Service Function. Password dialog box opens as shown
in Figure 4-1.
Figure 4-1 Password dialog box
Step 5. Enter the password kid in the Password box and then click "OK" button.
Step 6. Click Exit in Service Functions Menu. Then the E5071C exit the Measurement
View, then windows desktop screen appears with 3 icons (Network Analyzer,
Recycle Bin, and Internet Explorer.)
a. To get “Start” menu bar displayed, move the pointer to the bottom of the
screen with mouse.
b. Click “Start” and “Turn off Computer” in the pull down menu. “Turn off
Computer” dialog box opens.
c. Click “Turn off” button in the dialog box.
If you wish to return to Measurement View, double-click “Network
Analyzer” icon.
If you need to shut down the E5071C and again turn on, perform in
accordance with the following procedure.