240 Keysight E5071C ENA Network Analyzers
Replacement Procedure
Synthesizer Module Replacement (A1)
6-
Synthesizer Module Replacement (A1)
Tools Required
TORX screwdriver, T10, T15, and T20
•Flat edge screwdriver
Open-end wrench, 5/8 inch
Open-end torque wrench, 5/16 inch or 8mm (set to 1.08 N-m / 9.5 lb-in)
(for reconnecting SMA connector)
Torque driver set, 5/16 inch (set to 1.08 N-m / 9.5 lb-in) (for reconnecting
SMA connector)
Removal Procedure
Refer to Figure 6-25 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 203.
Step 2. Remove the front panel as described in “Front Panel Removal” on page 204.
Step 3. Disconnect the semi-rigid cable (item 1).
Step 4. Disconnect the semi-rigid cable (item 2).
Step 5. Remove the two TORX T10 screws fastening the module to the bottom. (item
3)
Step 6. Slide the synthesizer module (item 4).