26 Keysight E5071C ENA Network Analyzers
General Information
Required Equipment
1-
Adapter N(m)-3.5mm(f) Keysight p/n
1250-2879
2P
Torque
Wrench
Size: 20 mm
Torque: 90 N-cm
Keysight p/n
8710-1764
1P,A
Personal
Computer with
GPIB board
Windows XP, VEE7.5 or later 1 P,A
1. P: Performance Tests, A: Adjustment, T: Troubleshooting
2. The accuracy of a standard E4419A/B of 0.02 dB is adequate for all tests/adjust-
ment except “Dynamic Accuracy Test”. This test requires a power meter with Option
G12 or H12 that has been certified to a higher accuracy specification. If an Option
G12 or H12 power meter is not available, a test is provided on
“Power Meter
Accuracy Test” on page 31
to verify the accuracy of a standard power meter.
Table 1-3 Alternative Test Equipment
Equipment Critical specifications Model Qty
Use
1
Frequency
Counter
Frequency: 50MHz to 8.5GHz
Accuracy: < 2.5 ppm
Keysight 53131/2A
with Opt.010 and 124
1P,A
Multimeter Keysight 34401A
Keysight 34410A
Keysight 34411A
1
1
1
P, T
P, T
P, T
Function
Generator
Keysight 33220A
Keysight 33250A
1
1
P, T
P, T
Power Sensor No Substitute Keysight E9304A with
Opt. H19
1P,A
Short Type-N(m) Keysight 11512A 4 P,T
Load Type-N(m) Keysight 909F
Opt. 012
4P
1. P: Performance Tests, A: Adjustment, T: Troubleshooting
For the latest ETE lists, refer to Keysight Test Management Environment
(TME) Software Portal. The complete ETE lists are available in E5071C
Calibration Application WebHelp. http://cal.software.keysight.com/
Table 1-2 Recommended Test Equipment
Equipment Critical specifications Recommended Model Qty
Use
1