336 Keysight E5071C Service Manual
Firmware Update
Firmware Update
-
Step 2. Run the “E5071C_xx.xx.exe” to extract the E5071C.msi. Then, save this file to
a USB mass storage device.
Step 3. Connect the mouse and the keyboard to the E5071C rear panel. Then turn the
E5071C on.
Step 4. Connect the USB mass storage device into the front USB port of the E5071C.
Step 5. Update the E5071C firmware.
1. Press key.
2. Press Service Menu - Update Firmware on the menu bar, and the
Password dialog box will appear.
3. Enter the password e507xa into the Password box, then click OK.
Figure C-1 Password box
4. Open dialog box appears. Confirm that E5071.msi is displayed at the File
name column. Select it, then press the Open button, and the Windows
installer will appear.
Figure C-2 Firmware Open dialog box
Step 6. After the installation is finished, the instrument will be restarted and the
measurement view will appear.
Step 7. Press -Firmware Revision. Confirm the firmware revision you have just
installed.