352 Keysight E5071C Service Manual
Messages
Error Messages
-
73 Power meter not settled
The measurement value of the power meter is unstable, or it goes outside the
tolerance.
This error occurs when the power meter’s measurement value does not
become stable within about 10 seconds during measurement for power
calibration data or that for scalar-mixer calibration data.
This error also occurs when the power calibration value exceeds the set
tolerance.
In addition, it occurs when the power meter fails to respond for any reason,
such as power-down for example, during measurement under the same
conditions.
Ensure a firm connection of the power sensor.
241 Power on test failed
This error occurs when the power-on test fails, indicating a failure of this
instrument. Contact an Keysight Technologies sales office or the company
from which you bought the instrument.
61 Power unleveled
The out power level exceeds available range.
For example, if the level after correction exceeds the power level that can be
outputted when correcting the power level with the power calibration and
power slope feature, this error occurs.
Check that the power level is set correctly, the power calibration has been
performed correctly and the correction value of the power slope is set
correctly.
120 Printer error
This error occurs when the previous printing is still in progress or the printer
fails (offline, short of paper, etc.) at time of outputting the display image on the
LCD screen to the printer (HCOP:IMM command).
121 Print failed
This error occurs when printing fails for reasons other than Error 120, Printer
error.
-284 Program currently running
This error occurs when the PROG:SEL:STAT RUN command is executed with
the VBA program in the Run state.
-112 Program mnemonic too long
The length of the header exceeds 12 characters (see 7.6.1.4.1, IEEE488.2).