78 Keysight E5071C ENA Network Analyzers
Troubleshooting
Troubleshooting Using Diagnostics Test
4-
x: With
: Without
x: With
: Without
7 x x x x x A2-3 not found
8 x x x x x A2-3 not found
9 x x x x x A2-3 not found
10 x x x x x A2-3 not found
11 x x x x A2-2 not found
12 x x x x A2-2 not found
13 x x x x A2-2 not found
14 x x x x A2-2 not found
15 x x x A2-1 not found
16 x x x x x x A1-2 not found
17 x x x x x x A1-2 not found
18 x x x x x A1-1 not found
19 x x x x x x A3 not found
Table 4-2 Logic check for 4-ports E5071C
Attempt RCV1 RCV2 RCV3 RCV4 Synth-Source Synth-Local A83 Level
Vernier
Corresponding error
message
Table 4-3 Logic check for 2-ports E5071C
Attempt RCV1 RCV2 Synth-Source Synth-Local A83 Level
Vernier
Corresponding error
message
1 x x x x A2-2 not found
2 x x x x A2-2 not found
3 x x x A2-1 not found
16 x x x x A1-2 not found
17 x x x x A1-2 not found
18 x x x A1-1 not found
19 x x x x A3 not found