EasyManua.ls Logo

Keysight MP4300 Series - Page 141

Keysight MP4300 Series
198 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SYSTem:REBoot
Reboots the instrument to its power-on state.
Parameter Typical Return
(none) (none)
Reboots the instrument: SYST:REB
SYSTem:TIME <hh>, <mm>, <ss>
SYSTem:TIME?
Sets the time of the system clock. Specify hours (0 to 23), minutes (0 to 59), and seconds (0 to 59). The
real time clock is only used in conjunction with the BenchVue Power Control and Analysis software.
Parameter Typical Return
<hh>,<mm>,<ss> <hh,mm,ss>
Set the clock to 8:30 PM: SYST:TIME 20,30,0
SYSTem:VERSion?
Returns the SCPI version that the instrument complies with. Cannot be determined from front panel.
Parameter Typical Return
(none) <"version">
Return the SCPI version: SYST:VERS?
l
The command returns a string in the form "YYYY.V", where YYYY represents the year of the version
and V represents a version for that year.
Keysight MP4300 Series Operating and Service Guide 141
5 SCPI Programming Reference

Table of Contents

Related product manuals