[SOURce:]
CURRent
[:LEVel]
[:IMMediate]
[:AMPLitude] <value>, (@<chanlist>) Sets the output current when in current priority mode.
:LIMit
[:POSitive]
[:IMMediate]
[:AMPLitude] <value>, (@<chanlist>) Sets the positive current limit when in voltage priority mode.
:NEGative
[:IMMediate]
[:AMPLitude] <value>, (@<chanlist>) Sets the negative current limit when in voltage priority mode.
:DTABle
[:IMMediate]? (@<chanlist>) Returns the table currently being used by the channel.
:SASimulator
[:IMMediate]? (@<chanlist>) Returns the DAC table based on the SAS parameter settings.
:IMP? (@<chanlist>) Returns the calculated Imp of the internal DAC table.
:ISC? (@<chanlist>) Returns the calculated Isc of the internal DAC table.
:TABLe (@<chanlist>)
[:IMMediate]? <"name">, (@<chanlist>) Returns the DAC table based on the SAS parameter table settings.
:IMP? <"name">, (@<chanlist>) Returns the calculated Imp of the internal DAC table.
:ISC? <"name">, (@<chanlist>) Returns the calculated Isc of the internal DAC table.
:MODE FIXed|SASimulator|TABLe, (@<chanlist>) Selects the output mode of the instrument.
:PROTection
[:LEVel] <value>, (@<chanlist>) Sets the over-current protection level in SAS mode.
:STATe 0|OFF|1|ON, (@<chanlist>) Enables or disables over-current protection.
:SASimulator
:BWIDth "DEFAULT"|"DCDC_20UF"|"SHUNTSW"
(@<chanlist>)
Specifies the compensation in SAS mode.
:IMP <value>, (@<chanlist>) Sets the current at the Imp point of the curve in SAS mode.
:ISC <value>, (@<chanlist>) Sets the Isc current of the curve in SAS mode.
:SCALe <percent>, (@<chanlist>) Specifies the current scale in SAS mode.
:TABLe
:NAME <name, (@<chanlist> Activates a table when operating in Table mode.
[SOURce:]
DIGital
:INPut
:DATA? Reads the state of the digital control port.
:OUTPut
:DATA <value> Sets the state of the digital control port.
:PIN<1-7>
:FUNCtion <function> Sets pin function. DIO|DINPut|FAULt|INHibit|ONCouple|OFFCouple
:POLarity POSitive|NEGative Sets the pin polarity.
1 Quick Reference
24 Keysight MP4300 Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight MP4300 Series and is the answer not in the manual?

Keysight MP4300 Series Specifications

General IconGeneral
BrandKeysight
ModelMP4300 Series
CategoryPower Supply
LanguageEnglish

Related product manuals