constants are changed as indicated below. For current readback accuracy and temperature coefficient
of current readback, use the minus current specifications:
Current: All parallel specifications referring to current are twice the single output specification except
for programming resolution, which is the same for both single output and parallel output operation.
Voltage: All parallel specifications referring to voltage are the same as for a single output except for CV
load effect, CV load cross regulation, CV source effect, and CV short term drift. These are all twice the
voltage programming accuracy (including the percentage portion) at all operating points.