EasyManuals Logo

Keysight N9010B Instruction Manual

Keysight N9010B
2530 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1808 background imageLoading...
Page #1808 background image
4 System
4.6 Alignments
"Align Now All" on page 1808
"Align Now All but RF" on page 1810
"Align Now RF" on page 1812
"Align Now Expired" on page 1813
"Align Now Preselector" on page 1814
"Align Now All but RF Preselector" on page 1815
"Align Now RF Presel Only (20 Hz to 3.6 GHz)" on page 1815
"Align Now External Mixer" on page 1816
"Align Source" on page 1817
"Align Receiver" on page 1818
"Align Fast" on page 1818
"Align LO Leakage" on page 1819
"Align IF Cable" on page 1819
"Align RRH Amplitude" on page 1819
"Align LO Clock" on page 1820
"Align VXT Transceiver" on page 1821
"Align External Mixer Path" on page 1825
"Align Low Band" on page 1826
"Align High Band" on page 1826
4.6.2.1 Align Now All
In PXE, the key label is Align Now All (plus RF Presel 20 Hz 3.6 GHz)
Immediately executes an alignment of all subsystems. The instrument stops any
measurement currently underway, performs the alignment, then restarts the
measurement from the beginning (similar to pressing the Restart key).
If an interfering user signal is present at the RF Input, the alignment is performed on
all subsystems except the RF. After completion, the Error Condition message “Align
RF skipped” is generated. In addition, the Error Condition message “Align Now, RF
Short Range Comms & IoT Mode User's &Programmer's Reference 1808

Table of Contents

Other manuals for Keysight N9010B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9010B and is the answer not in the manual?

Keysight N9010B Specifications

General IconGeneral
BrandKeysight
ModelN9010B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals