8 Trigger
8.1 Trigger
8.1.1.15 Q (Demodulated)
When Q (Demodulated) is selected, the trigger condition is met when the Q voltage
crosses the Q voltage trigger level.
This trigger type is only valid for measurements that support the I/Q inputs.
The Trigger tab contains the following Trigger Source dependent controls when Q
(Demodulated) Trigger is selected:
"Prot Frame Aligned" on page 2340
"Trigger Delay" on page 2343
"Trigger Slope" on page 2347
Additional controls are also present that are not dependent on the selected Trigger
Source.
Example
:TRIG:<meas>:SOUR QDEM
Annunciation Q (Demod) (in the Meas Bar)
8.1.1.16 Aux I/Q Mag
When Aux I/Q Mag is selected, the trigger condition is met when the auxiliary
receiver's I/Q magnitude output crosses the Auxiliary I/Q magnitude trigger level.
This trigger type is only valid for measurements that support the I/Q inputs.
The Trigger tab contains the following Trigger Source dependent controls when Aux
I/Q Mag Trigger is selected:
"Prot Frame Aligned" on page 2340
"Trigger Delay" on page 2343
"Trigger Slope" on page 2347
"Trigger Center Frequency" on page 2357
"Trigger BW" on page 2357
Additional controls are also present that are not dependent on the selected Trigger
Source.
Example
:TRIG:<meas>:SOUR AIQM
Annunciation Aux I/Q Mag (in the Meas Bar)
Short Range Comms & IoT Mode User's &Programmer's Reference 2338