EasyManua.ls Logo

Keysight N9010B - FCS Type; Decoding Level; 3.4.8.3 Meas Time; Search Length

Keysight N9010B
2530 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Short-Range Comms & IoT Mode
3.4 HRP UWB Demodulation Measurement
FCS Type
The MAC may optionally employ the 4-octet FCS with the HRP UWB PHY in HPRF
mode, but in all other HRP UWB PHY modes shall employ the 2-octet FCS.
Remote
Command
[:SENSe]:HUWB:DATA:FCSType OCTet2 | OCTet4
[:SENSe]:HUWB:DATA:FCSType?
Example
:HUWB:DATA:FCST OCT4
Couplings Can be set to 4-octet when "PHY Mode" on page 501 is ERDEV-HPRF, while in other PHY modes, this
value can be only set to 2-octet. This parameter is disabled when STS packet config is SP3 (After SFD
No Data).
Preset
OCT2
State Saved Saved in instrument state
Range
2-octet | 4-octet
Decoding Level
Specify decoding level.
Remote
Command
[:SENSe]:HUWB:DATA:DECoding NODecoding | FEC | PSDU
[:SENSe]:HUWB:DATA:DECoding?
Example
:HUWB:DATA:DEC FEC
Couplings This parameter is not available when STS packet config is SP3 (After SFD No Data) or Data Analysis is
off.
Preset
No Decoding
State Saved Saved in instrument state
Range
No Decoding | PHR+PSDU bits with FEC | PSDU
3.4.8.3 Meas Time
Sets time-related parameters for the demodulation.
Search Length
Search Length specifies the amount or length of the Search Time input data that is
included in the measurement pulse search.
Remote Command
[:SENSe]:HUWB:TIME:SLENgth <time>
[:SENSe]:HUWB:TIME:SLENgth?
Short Range Comms & IoT Mode User's &Programmer's Reference 513