EasyManua.ls Logo

Keysight N9010B - 6.10.6 Calibrator Reference

Keysight N9010B
2530 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Input/Output
6.10 Calibrator Control
Preset Unaffected by Mode Preset. Set to 0Hz by Restore Input/Output Defaults or Restore System
Defaults->All
Min/Max Dependent on the selected calibrator
6.10.6 Calibrator Reference
Determines the frequency reference type used by the RCal module of the currently
selected Cal Group
Remote Command
:SYSTem:CALibration:TUNE:REFerence INTernal | EXTernal
:SYSTem:CALibration:TUNE:REFerence?
Example Set the calibrator frequency reference to Internal:
:SYSTem:CALibration:TUNE:REFerence INTERNAL
Dependencies Only displayed when an RCal module is the selected calibrator
Preset Unaffected by Mode Preset. Set to preset value by Restore Input/Output Defaults or Restore System
Defaults->All
Range
INTernal|EXTernal
2217 Short Range Comms & IoT Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight N9010B

Related product manuals