EasyManua.ls Logo

Keysight N9010B - Negative Offset Limit(Remote Command only)

Keysight N9010B
2530 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Short-Range Comms & IoT Mode
3.7 ACP Measurement
Modes Values
SA -45, -60, 0, 0, 0, 0 | -45, -60, 0, 0, 0, 0
WCDMA -44.2, -49.2, -49.2, -49.2, -49.2, -49.2 | -32.2, -42.2, -42.2, -42.2, -
42.2, -42.2
LTEAFDD, LTEATDD,
MSR
-44.2,-44.2,-44.2,-44.2,-44.2,-44.2 | -29.2, -29.2, -29.2, -29.2, -29.2,
-29.2
5G NR -43.8, -43.8, -43.8, -43.8, -43.8, -43.8 | -29.2, -29.2, -29.2, -29.2, -
29.2, -29.2
Radio Test -60, -60, -60, 0, 0, 0
State Saved Saved in instrument state
Min/Max -150.0/50.0
Backwards
Compatibility
SCPI
:CALCulate:MCPower:OFFSet:LIST:LIMit:POSitive[:UPPer]:DATA
(Power Suite)
Negative Offset Limit(Remote Command only)
Enables you to set the upper limit for the lower segment of the specified offset pair.
Remote
Command
:CALCulate:ACPower:OFFSet[1]|2[:OUTer]:LIST:LIMit:NEGative[:UPPer]:DATA
<real>,...
:CALCulate:ACPower:OFFSet[1]|2[:OUTer]:LIST:LIMit:NEGative[:UPPer]:DATA?
Subopcode: 1 = BTS/Downlink (Default), 2 = MS/Uplink
Example
:CALC:ACP:OFFS:LIST:LIM:NEG:DATA 0,0,0,0,0,0
:CALC:ACP:OFFS:LIST:LIM:NEG:DATA?
Notes Some Modes do not support Offset subopcode 2. In those cases, commands with subopcode 2 are
accepted without error but ignored
Preset When "Max Num of Offsets" on page 902 is 12, the preset value of Offset G ~ L is the same as the Offset
F value
Modes Values
SA -45, -60, 0, 0, 0, 0 | -45, -60, 0, 0, 0, 0
WCDMA -44.2, -49.2, -49.2, -49.2, -49.2, -49.2 | -32.2, -42.2, -42.2, -42.2, -
42.2, -42.2
LTEAFDD, LTEATDD,
MSR
-44.2, -44.2, -44.2, -44.2, -44.2, -44.2 | -29.2, -29.2, -29.2, -29.2, -
29.2, -29.2
5G NR -43.8, -43.8, -43.8, -43.8, -43.8, -43.8 | -29.2, -29.2, -29.2, -29.2, -
29.2, -29.2
Radio Test -60, -60, -60, 0, 0, 0
Short Range Comms & IoT Mode User's &Programmer's Reference 885

Table of Contents

Other manuals for Keysight N9010B

Related product manuals