3 Short-Range Comms & IoT Mode
3.3 LoRa (CSS) Demodulation Measurement
Remote Command
[:SENSe]:LORA:FREQuency:SPAN <freq>
[:SENSe]:LORA:FREQuency:SPAN?
Example
:LORA:FREQ:SPAN 10 MHz
:LORA:FREQ:SPAN?
Dependencies If the electrical attenuator is enabled, any attempt to set the span such that the stop Frequency would
be greater than 3.6 GHz will result in an error
Preset 75 kHz
State Saved Saved in instrument state
Min 10 Hz
Max Hardware dependent:
–
No Option = 10 MHz
–
Option B25 = 25 MHz
–
Option B40 = 40 MHz
–
Option B85 = 85.0 MHz
–
Option B1A = 125.0 MHz
–
Option B1X = 140 MHz
–
Option B1Y = 160 MHz
–
Option B2X = 160 MHz
–
Option B5X = 160 MHz
–
M941xA-B3X = 160 MHz
–
M941xA-B6X = 160 MHz
M941xA-B12 = 160 MHz
Annotation Span <value> appears on the first line of the annotation in the lower right corner of the RF Spectrum
window
CF Step
Modifies the step size of the center frequency, the start frequency, and the stop
frequency. The step size is useful for finding the harmonics and the sidebands
beyond the current frequency span of the analyzer.
Remote Command
[:SENSe]:FREQuency:CENTer:STEP[:INCRement] <freq>
[:SENSe]:FREQuency:CENTer:STEP[:INCRement]?
Example
:FREQ:CENT:STEP 500 MHz
:FREQ:CENT UP
Short Range Comms & IoT Mode User's &Programmer's Reference 383