3 Short-Range Comms & IoT Mode
3.4 HRP UWB Demodulation Measurement
Range
ON|OFF
Transmit Mask Stitching Period
Specifies repeat period of input signal.
Remote
Command
[:SENSe]:HUWB:TRANsmit:MASK:STITching:PERiod <time>
[:SENSe]:HUWB:TRANsmit:MASK:STITching:PERiod?
Example
:HUWB:TRAN:MASK:STIT:PER 1ms
:HUWB:TRAN:MASK:STIT:PER?
Preset 2.909 ms
State Saved Yes
Min 1 usec
Max 100 msec
3.4.9 Recall
3.4.9.1 Signal Configuration
Selects a Signal Studio N7610C scp file from which to import the demod settings to
the HRP UWB measurement.
N7610C Setup Files:
Extension:scp
Signal Studio N7610C (IoT) configuration file, when PHY Specification is 802.15.4
HRP UWB.
Remote Command
:MMEMory:LOAD:HUWB:SETup <string>
Example
:MMEM:LOAD:HUWB:SET "mySetup.scp"
Dependencies Only appears in Short Range Comms Mode, in the HRP UWB measurement
Annotation After recall is complete, an advisory is displayed in the message bar confirming that the filter’s coeffi-
cients were loaded
Status Bits/OPC
dependencies
Sequential aborts the current measurement
3.4.10 Sweep
Accesses controls to configure and control the acquisition of data, and the X-axis
parameters of the instrument.
528 Short Range Comms & IoT Mode User's &Programmer's Reference

Table of Contents

Other manuals for Keysight N9010B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9010B and is the answer not in the manual?

Keysight N9010B Specifications

General IconGeneral
BrandKeysight
ModelN9010B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals