EasyManua.ls Logo

Keysight N9010B - Advanced Demod Setup; Phase Tracking; Initial Frequency Error Compensation

Keysight N9010B
2530 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Short-Range Comms & IoT Mode
3.4 HRP UWB Demodulation Measurement
Example
:HUWB:COMP:CCLO ON
:HUWB:COMP:CCLO?
Preset
ON
State Saved Yes
Range
OFF|ON
Advanced Demod Setup
Configures advanced demodulation parameters.
Phase Tracking
Enables phase tracking.
Remote Command
[:SENSe]:HUWB:PHASe:TRACking[:STATe] OFF | ON | 0 | 1
[:SENSe]:HUWB:PHASe:TRACking[:STATe]?
Example
:HUWB:PHAS:TRAC ON
:HUWB:PHAS:TRAC?
Preset
ON
State Saved Yes
Range
OFF|ON
Initial Frequency Error Compensation
Enables or disables initial frequency error estimation and correction. Initial
estimation has a lock range of about +/- 1 MHz.
Remote
Command
[:SENSe]:HUWB:COMPensation:FREQuency:ESTimation FRAMe | SYNC
[:SENSe]:HUWB:COMPensation:FREQuency:ESTimation?
[:SENSe]:HUWB:COMPensation:FREQuency[:STATe] OFF | ON | 0 | 1
[:SENSe]:HUWB:COMPensation:FREQuency[:STATe]?
Example
:HUWB:COMP:FREQ:EST SYNC
:HUWB:COMP:FREQ:EST?
:HUWB:COMP:FREQ ON
:HUWB:COMP:FREQ?
Preset ON
State Saved Yes
Yes
Range Est from Sync | Est from Entire Frame
516 Short Range Comms & IoT Mode User's &Programmer's Reference