3 Short-Range Comms & IoT Mode
3.4 HRP UWB Demodulation Measurement
:CALC:HUWB:LIM:TEST?
Preset
ON
State Saved Yes
Range
OFF|ON
Main Lobe Width
Sets the limit for main lobe width pass/fail test.
Remote Command
:CALCulate:HUWB:LIMit:MLWidth <time>
:CALCulate:HUWB:LIMit:MLWidth?
Example
:CALC:HUWB:LIM:MLW 1ns
:CALC:HUWB:LIM:MLW?
Preset 500ps
State Saved Yes
Min 0ns
Max 10ns
Side Lobe Peak
Sets the limit for side lobe peak pass/fail test.
Remote Command
:CALCulate:HUWB:LIMit:SLPeak <real>
:CALCulate:HUWB:LIMit:SLPeak?
Example
:CALC:HUWB:LIM:SLP 0.3
:CALC:HUWB:LIM:SLP?
Preset 0.3
State Saved Yes
Min 0.0
Max 1.0
Frequency Error
Sets the limit for frequency error pass/fail test.
Remote Command
:CALCulate:HUWB:LIMit:FERRor <real>
:CALCulate:HUWB:LIMit:FERRor?
Example
:CALC:HUWB:LIM:FERR 20
:CALC:HUWB:LIM:FERR?
518 Short Range Comms & IoT Mode User's &Programmer's Reference