4 System
4.6 Alignments
Query Temperature of Last Align Now External Mixer Path
Remote Command
:CALibration:TEMPerature:INTernal:EMPath?
Example
:CAL:TEMP:INT:EMP?
Notes Value in degrees Centigrade when the last successful Align Now External Mixer Path was executed
Dependencies Only appears option MXP is installed
State Saved No
Query Time since Last Align Now Low Band
Remote Command
:CALibration:TIME:INTernal:LBANd?
Example
:CAL:TIME:INT:LBAN?
Notes Value in hours since the last successful Align Now Low Band
Dependencies Only appears in VXT models M9415A/16A
State Saved No
Query Temperature of Last Align Now Low Band
Remote Command
:CALibration:TEMPerature:INTernal:LBANd?
Example
:CAL:TEMP:INT:LBAN?
Notes Value in degrees Centigrade when the last successful Align Now Low Band was executed
Dependencies Only appears in VXT models M9415A/16A
State Saved No
Query Time since Last Align Now High Band
Remote Command
:CALibration:TIME:INTernal:HBAN?
Example
:CAL:TIME:INT:HBAN?
Notes Value in hours since the last successful Align Now High Band
Dependencies Only appears in VXT models M9415A/16A
State Saved No
Query Temperature of Last Align Now High Band
Remote Command
:CALibration:TEMPerature:INTernal:HBANd?
Example
:CAL:TEMP:INT:HBAN?
Notes Value in degrees Centigrade when the last successful Align Now High Band was executed
Dependencies Only appears in VXT models M9415A/16A
State Saved No