3 Short-Range Comms & IoT Mode
3.7 ACP Measurement
n Results Returned
# Measurement Type Item Unit, if any
4
TPRef
Relative power or dBc
PSDRef
Relative PSD dB
5
TPRef
Absolute power dBm
PSDRef
Absolute PSD dBm/Hz, dBm/MHz*
6
TPRef
Reference power dBm
PSDRef
Reference PSD dBm/Hz, dBm/MHz*
7 Reference Index 1
8 Reference Index 2
9 0 (Reserved)
10 0 (Reserved)
10(f - 1) + 1 Inner = 1 or Outer = 2
10(f - 1) + 2 Offset A~L. (A = 1, B = 2, ... L = 12)
10(f - 1) + 3 Offset Side. Lower = 1 or Upper = 2
10(f - 1) + 4
TPRef
Relative power dBc
PSDRef
Relative PSD dB
10(f - 1) + 5
TPRef
Absolute power dBm
PSDRef
Absolute PSD dBm/Hz, dBm/MHz*
10(f - 1) + 6
TPRef
Reference power dBm
PSDRef
Reference PSD dBm/Hz, dBm/MHz*
10(f - 1) + 7 Reference Index 1
10(f - 1) + 8 Reference Index 2
10(f - 1) + 9 0 (Reserved)
10(f - 1) + 10 0 (Reserved)
"Measurement Type" on page 947 determines which type of power result is returned: Total Pwr Ref
(TPRef) or PSD Ref (PSDRef)
*For PSD results, the unit is determined by "PSD Unit" on page 955: DBMHZ or DBMMHZ
If any result is not available, 9.91E+37 (NaN) is returned
11 Returns Marker Table data as a series of comma separated values in the following form:
<Marker Number>,<Marker Trace>,<X>,<Y>,<Reserved>,<Reserved>
Only markers that are enabled are included. <Reserved> are returned as NaN (“Not a Number”, 9.91e+37).
The data is returned in the current sort order as displayed in the Marker Table
Short Range Comms & IoT Mode User's &Programmer's Reference 775

Table of Contents