DIAGNOSTICS Checkout Procedures
10DEC05
DG4825-1
Page
144 of 180
3 Measure the following voltages at the indicated test points.
4 If any of the voltages do not match, install a new DIGITIZER BOARD A3.
Voltage
MULTIMETER
Ground Positive
+5 ± 0.1 V DC TP2 TP3
+24 ± 2.5 V DC TP2 TP1

Table of Contents