DIAGNOSTICS Checkout Procedures
10DEC05
DG4825-1
Page
153 of 180
4 Touch:
[LASER ON]
[READ]
5 Check:
“Laser Power” is 28 - 32 mW
“Percent of original power” is 90%
6 Are the values correct?
Yes No
Continue with the next test. 1. Touch [SWEEP].
2. Check that the laser beam scans.
3. Check the connections to the LASER
DRIVER BOARD A17.
4. Install a new OPTICS MODULE.