KurzBSeriesOperationsGuide 6–3
Flow Meter Configuration
Modbus Protocol Write Coil Command
Bywritinga1tothespecificModbuscoil,youcantriggerazero(#0),mid(#1),span(#2),and
cycleall(#3,function05)driftcheck.OnlyoneModbusdriftcheckstartcommandcanrunata
time.ThistriggerdoesnotrequiretheAutoDriftChecktobeON.
However,iftheAutoDriftCheck
isONandcurrentlyperformingadriftcheck,theModbustriggereddriftcheckwillnotstartuntil
theotherdriftcheckcompletes.
Modbus Status Registers
StatusregistersareprovidedviaModbus(function02)toindicatewhetheranyofthegiven
optionsareactive(1)orinactive(0).
Table 61. ModbusWriteCoilCommands ‐Function0x05WriteSingleCoil,
Function0x01ReadCoils
Coil# Description Hex
0Startzerodriftcheck. 000
1Startmidspandriftcheck. 001
2
Startspandriftcheck.
002
3Startdriftcheckcycle. 003
4Abortongoingdriftcheck. 004
5Reserved. 005
6SaveworkingconfigurationtoEEPROM 006
7Reserved. 007
8Startpurge(requiresPurgeoption). 008
9Resetdevicetotalizer 009
Table 62. ModbusStatusRegistersFunction0x02ReadDiscreteInputsOnly
Coil# Description Hex
0Zerodriftcheckstarted. 000
1Midspancheckstarted. 001
2Spancheckstarted. 002
3Driftcheckcyclestarted. 003
47Reserved. 004007
8Purgecyclestarted. 008
915 Reserved. 009
16 RPresistanceabovehighlimit(error). 010
17 RPresistancebelowlowlimit(error). 011

Table of Contents

Other manuals for Kurz B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kurz B Series and is the answer not in the manual?

Kurz B Series Specifications

General IconGeneral
BrandKurz
ModelB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals