KurzB‐SeriesOperationsGuide2–28
B‐SeriesFlowMeterDisplay Mode Options
Option 44 — Input Voltage
Note ThisOptionisusedbyKurzfordiagnosticandtroubleshootingpurposes.
ToviewtheInputVoltagemenuinDisplaymode:
1> PressD.
2> Press 2toinvoketheQuickJumpoption.
3> Press44fortheInputVoltagemenu,andthenpressE.
VPs= 2.145504
VIph= 0.083602
Thepromptshowingsensorcircuitvoltagesappears.
4> PressPtocontinue.
VLI= 0.088409
VLS= 3.000270
Thepromptshowingsensorcircuitvoltagesappears.
5> PressPtocontinue.
VRtch= 2.263744
VRtc1= 0.009368
Thepromptshowingsensorcircuitvoltagesappears.
6> PressPtocontinue.
VExt= 0.002873
VTemp= 1.432526
Thepromptshowingthevoltagefromtheexternalinputsignalandthevoltageto
determinethesensorcontrolboardtemperatureappears.
7> PressPtocontinue.
VCal= 0.000984
Thepromptshowingvoltagemeasuredfr omthesensorcontrolboardvoltagesource
appears.
8> PressHtoexit.
ThemainDisplaymode(DSP)promptappears.