KurzBSeriesOperationsGuide4–24
Program ModeAdvanced Setup
Option 10 — Setting Up the NE-43 Alarm
TheflowmetersupportstheNAMURspecificationNE43alarmonthe420mAsignal.Therange
forthenonalarmedanalogoutputsignalisbetween3.8and20.5mA.

Figure 41. 420mAoperationwiththeNE43alarm
ThefollowingflowmetererroreventscauseaNE43alarm:
Externalanaloginputoutofrange
Sensortypedoesnotmatchboardbuild
UnabletowriteconfigurationfiletoEEPROM
Abnormalsensornodevoltages
Sensortypedoesnotmatchconfiguration
Sensorovervoltagecrowbarengaged
Sensorcontroldrivestoppedresponding
ADCfailedtoconvertmeasurement
Highsensororwireleakage
Rpssensorleadopencircuit
Wireloopresistanceabovehighlimit
Rtcresistancebelowlowlimitandabovehighlimit
Rpresistancebelowlowlimitandabovehighlimit
Flowrateabovedesignlimit
ToaccesstheNE43AlarmSetupmenuinProgrammode:
1> PressP.
2> EntertheAdvancedaccesscode,andthenpressE.
3> Press 2toinvoketheQuickJumpoption.
4> Press10fortheNE43AlarmSetupmenu,andthenpressE.
NE-43 ALARM TYPE
>HIGH OUTPUT ^v
TheNE43alarmtypepromptappears.
ALOWOUTPUTdrivesthe420mAoutputto3.6mAwhenameterfaultoccurs.
AHIGHOUTPUTdrivesthe420mAoutputto21mAwhenameterfaultoccurs.
NE-43 Alarm
(low value kick-out)
NE-43 Alarm
(high value kick-out)
Saturate Saturate
Normal Operation
3.6 3.8 20.5 210 mA

Table of Contents

Other manuals for Kurz B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Kurz B Series and is the answer not in the manual?

Kurz B Series Specifications

General IconGeneral
BrandKurz
ModelB Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals