KurzB‐SeriesOperationsGuide 2–29
B‐SeriesFlowMeterDisplay Mode Options
Option 45 — Sensor Output
Note ThisOptionisusedbyKurzfordiagnosticandtroubleshootingpurposes.
ToviewtheSensorOutputmenuinDisplaymode:
1> PressD.
2> Press 2toinvoketheQuickJumpoption.
3> Press45fortheSensorOutputmenu,andthenpressE.
IRp= 0.1656 A
PRp= 0.3089 W
Thepromptshowingsensordrivecalibrationinformationappears.
4> PressPtocontinue.
Rp= 12.0637 OHM
Rtc= 335.236 OHM
Thepromptshowingsensorresistanceappears.
5> PressPtocontinue.
TRp= 195.2 DEGF
TRtc= 87.6 DEGF
Thepromptshowingthetemperaturemeasuredbyeachsensorappears.
6> PressPtocontinue.
RLl= 0.1088 OHM
Thepromptshowingsensorloopwireresistanceappears.
7> PressPtocontinue.
SMPS= 0.02
SFPM= 8.86
ThepromptshowingtheflowmeasurementinEnglishandmetricunitsappears.
8> PressHtoexit.
ThemainDisplaymode(DSP)promptappears.