KurzB‐SeriesOperationsGuide2–16
B‐SeriesFlowMeterDisplay Mode Options
Option 15 — Variable Flow Correction Data
Thisisusedtocorrecttheflowcalibrationdatatoaccountforflowdisturbancesovertheentire
velocityrangeoftheflowmeter.Therecanbeuptoeightcorrectionpointsonadatacurve.
ToviewtheVariableFlowCorrectionDatamenuinDisplaymode:
1> PressD.
2> Press 2toinvoketheQuickJumpoption.
3> Press15fortheVariableFlowCorrectionDatamenu,andthenpressE.
ENTER # OF FLOW
DATA SET >1
Thepromptforthedatapairappears.
4> Enterthedatapairnumber(1‐8)youwanttomodify/enter.PressPtocontinue.
ENTER FLOW DATA
SET # >1
Thepromptforthefirstsetappears.
5> PressPtocontinue.
REF VALUE R1
>1.200000000 SMPS
TheR1valueforSet1appears.
6> PressPtocontinue.
METER DATA D1
>0.66700000 SMPS
TheD1valueforSet1appears.
7> PressPtocontinuetoscrollingthroughallthesets,orpressHtoexit.
ThemainDisplaymode(DSP)promptappears.