KurzB‐SeriesOperationsGuide 1–17
B‐SeriesFlowMeterModes & Menus
• Remotecorrectionfactorstatus(enabledordisabled)
• Systemunit(English/metric)
Run Data Log
TheRunDatalogprovidesthefollowinginformation:
• Currentruntimeinseconds
• Devicetagname
• Snapshotofthemeasuredflowrate,totalizedflow,andtemperature
Extended Utilities Mode
ExtendedUtili tiesmodeallowsaccesstoadvancedmeterfunctionsforinitiatingdriftchecksand
resettingtheflowtotalizer.Thisisanaccesscode‐protectedmode.Thedefaultaccesscodeis
654321.
ToenterExtendedUtilitiesmode:
1> PressE.
2> Entertheaccesscode.
3> PressE.
4> YoucanscrollthroughthemenuorquickjumptoanyofthefunctionsshowninTable 1‐3.
Note Functionswith()canbeexecutedviaHartcommand.
Table 1‐3. ExtendedUtilitiesModeFunctions
Option# Function Description/Parameters
1ZERODRIFT DriftCheckatZero()
IfYES,runstestandprovidesupdated%Diff,VIN,and VOUTdata.
IfNO,showslast%Diff,VIN,andVOUTdata.
2MIDDRIFT DriftCheckatMid‐span()
IfYES,runstestandprovidesupdated%Diff,VIN,
andVOUTdata.
IfNO,showslast%Diff,VIN,andVOUTdata.
3SPANDRFT DriftCheckatFull‐span()
IfYES,runstestandprovidesupdated%Diff,VIN,and VOUTdata.
IfNO,showslast%Diff,VIN,andVOUTdata.
4CYCLEDRIFT DriftCheckCycle(alltests)
()
IfYES,runstestandprovidesupdated%Diff,VIN,and VOUTdata.
IfNO,showslast%Diff,VIN,andVOUTdata.
5RESETTOTAL ResetFlowTotal()
IfYES,resetscounterstozero.
IfNO,returnstothemenu.