KurzBSeriesOperationsGuide2–14
BSeriesFlowMeterDisplay Mode Options
GAS NAME
>Air
Thegasnamepromptappears.
Gasnamescanbereferredtobytheirmolecularformulaorcommonname.Only16
charactersappearinthedisplay,butupto40charactersarestoredinternally.

9> PressPtocontinue.
GAS MOL WT
>28.9600000
Themolarweightpromptappears.
10> PressPtocontinue.
NEW REF DENSITY:
1.9684 KG/M3
Thereferencedensityisautomaticallycalculatedbasedonthevaluesenteredforthe
referencetemperatureandreferencepressure.
11> PressPtocontinue.
Gas
Molecular
Formula
Molar
Weight
Gas
Molecular
Formula
Molar
Weight
Air Air 28.97 Helium He 4.00
Argon Ar 39.94 Hydrogen H2 2.02
Butane C4H10 58.12 Digestergas
(50CH4,50CO2)
Carbondioxide CO2 44.01 Digestergas
(60CH4,40CO2)
Dryammonia NH3 17.04 Digestergas
(70CH4,30CO2)
Drychlorine CL2 71.0 Nitrogen N2 28.02
Ethane C2H6 30.07 Oxygen O2 32.00
Ethylene C2H4 28.05 Propane C3H8 44.09