EasyManua.ls Logo

LABORIE Flowmaster - Page 91

LABORIE Flowmaster
95 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Electro Magnetic Compatibility 91
electronic equipment.
RF emissions
CISPR 11
Class A
The EMISSIONS characteristics of the
Flowmaster make it suitable for use in
industrial areas and hospitals.
(CISPR 11 class A). If it is used in a
residential environment (for which
CISPR 11 class B is normally required)
this
equipment might not offer adequate
protection to radio-frequency
communication services. The user might
need to take mitigation measures, such
as relocating or re-orienting the
equipment.
Harmonic
emissions IEC
61000-3-2
Class A
Voltage
fluctuations/flicker
emissions IEC
61000-3-3
Complies
WARNING
When the Flowmaster is used adjacent to or stacked with
other equipment its normal operation should be verified.
Guidance and manufacturer’s declaration electromagnetic immunity
The Flowmaster is intended for use in the electromagnetic environment specified
below. The customer or the user of the Flowmaster should assure that it is used
in such an environment.
When the Flowmaster system is exposed to electromagnetic (EM) disturbances,
the system may show abnormal behavior. For instance, the measured traces
may be become obscured, the software may crash or, in case of very high level
transient voltage events like for instance ESD, parts of the system may even
become defective.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
 kV contact
15 kV air
 kV contact
15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.

Table of Contents

Related product manuals