38 Laborie Flowmaster User’s and Service Manual | LBL-000990 [B] Chapter 5
Time base
This function is available when the print orientation is portrait as set in the
previous setting. It allows you to show the complete investigation in one graph
(default), or split the investigation in different graphs of 15, 30, 60 or 120
seconds of time range.
Flow scale
The maximum value of the flow graph is default 25 ml/s.
You can also select 50 ml/s, 75 ml/s, 100 ml/s and Auto scaling.
Volume scale
The maximum value of the volume graph is default 500 ml.
You can also select 250 or 1000 ml.
Post flow time
When no flow detected for 60 seconds (default), the uroflow recording is
ended automatically. You can also select 15, 30, 60, 120, 180 or 600 seconds.
Online graph size
During the investigation (without using the database) default a medium sized
graph is displayed. You can select Off (no graphs) in which case no graph is
displayed, and select small sized graphs, medium sized graphs or large sized
graphs.
Required patient fields
Default last name and date of birth are required fields while entering a new
patient in the database. As required field, you can also select ‘patient number’,
‘personal number’ or ‘personal number and investigator’.
5.2.4 Flow Scheduler
Select Flow scheduler in the flow program to define for each day of the week
the times that the flowmeter must be on.