The configuration parameters are described below.
• N
AME: name of the MASTER SET to be programmed.
• D
ESCRIPTION: description of the MASTER SET (optional).
• T
YPE: MASTER SET TYPE
• ZEROING: one static master
• S
ENSITIVITY: two static masters
• Z
EROING & SENSITIVITY: three static masters
• D
YNAMIC ZEROING: a dynamic master
• P
RECISION: defines the number of places after the decimal point for the dimensions
that make up the M
ASTER SET (options: x1, x0.1, x0.01, x0.001, x0.0001, x0.00001).
• M
EASURE UNIT: measurement units for the dimensions that make up the MASTER
SET (options: mm, in, deg, g, °C, °F, mm).
3.1.2 Dimensions TAB
Afterwards tap on the DIMENSIONS tab where, according to the number of dimensions of
the M
ASTER SET, it is possible to set the ZERO LIMIT, the PERIODIC LIMIT and the ZERO MASTER
value for each dimension. If M
ASTER SET TYPE foresees up to three masters for zeroing
and sensitivity correction, two more master values must be defined for each dimension.
Tap on the field to be changed, and then E
DIT the value tapping on the relevant
button.
• ZERO LIMIT: this value is the maximum permissible difference between the last
acquired zero and the first.
• P
ERIODC LIMIT: this value is the maximum permissible difference between the
last acquired zero and the last but one.
• Z
ERO MASTER: this is the certified value of the master associated with the zero.