EasyManua.ls Logo

marposs WRP45 - Page 64

Default Icon
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WRS Installation Manual
64
In this case, the PROBE signal Status parameter when the ERROR signal is active (see para. 4.12.6 point 3)
must be set to “D” (deflected). This means that the probe is considered to be deflected when the ERROR
signal is active. The following diagram illustrates the TOUCH sub-routine, which is designed to prevent the
machine from being stopped unnecessarily when the ERROR signal is activated temporarily.
Figure 4-52. Touch sub-routine. Example application.
If the communication problem is persistent or too frequent the error will be activated permanently.
START
Save starting position
Touch movement
(TOUCH)
Error
signal
active?
RETURN
Error
signal
active?
NO
YES
Stop machine
Allarm
MSG (probe alarm signal)
Attesa 2 sec.
YES
NO
Return to start
position

Table of Contents