4.15.2.1 System pairing procedure
To replace or add a new probe, it is necessary to perform the
WRP - WRTS transmitter and WRI Receiver Pairing function. For
this aim follow the procedure described below: A WRP45 probe is
used in the procedure described, but the operations are the same
also if a WRP60 or WRTS probe is used.
Figure 4-60. Pairing procedure menu.
Insert the batteries into their compartment
respecting the polarity indicated on the printed
circuit.
Close the batteries’ compartment, and wait for the
first Red flash of the LED’s.
Access the programming menu of the receiver/interface by pressing PROGRAM on the remote control the
ENTER optical button covered for 5 seconds at least.