12-8 Probes and Biopsy
3. Secure the sheath with enclosed elastic
bands.
4. Inspect the sheath to ensure there are no
holes or tears.
Probe sizes
65EC10EB: (L) 173.06 mm, (D) 32.64 mm
D
L
The above is a dimension illustration of the intra-cavity probe. Customers can purchase
appropriate sterile sheaths according to the provided probe sizes. We recommend a general
kind of transducer cover produced by CIVCO which identified as GENERAL PURPOSE
LATEX TRANSDUCER COVERS (Sterile), 610-214(E8285CN,UA0006).
12.1.5 Probes Cleaning and Disinfection
After completing each examination, clean and disinfect the probes as required. When biopsy
procedures have been performed, be sure to sterilize the needle-guided bracket. Fail to do
so may result in the probe and the needle-guided bracket to becoming sources of infection.
Please follow the instructions in the manual for cleaning.
WARNING:
Never immerse the probe connector into liquid such as water or
disinfectant. Immersion may cause electrical shock or
malfunction.
CAUTION:
1.
When performing cleaning and
disinfection of the probe to prevent
infection, wear sterile gloves.
2.
After disinfection, rinse the probe thoroughly with sterile
water to remove all chemical residues. Chemical residues on
the probe may be harmful to the human body.
3
No cleaning and disinfecting may result in the probe
becoming a source of infection.
4.
Please follow the disinfectant manufacturer's manual for
performing cleaning and disinfection, including preparing
sterile water and cleaning and disinfection time.