Image Optimization 5-9
Operation
Select the colorize map through the [Colorize Map] item in the menu;
The system provides 16 different maps to be selected among.
Impacts
The function is available in real-time imaging, freeze or cine review status.
Gray Map
Description
Adjusting grayscale contras to optimize the image.
Operation
Select among the maps through the [Gray Map] item in the menu;
You can also adjust through the grayscale bar: move the cursor to the grayscale
bar and press <Set> on the control panel to adjust.
The system provides 8 gray maps to be selected among.
Impacts
The function is available in real-time imaging, freeze or cine review status.
Edge Enhance
Description
This function is used to increase image profile, so as to distinguish the image
boundary.
Operation
Adjust through the [Edge Enhance] item in the menu;
The system provides 14 levels of edge enhance effects, off represents no edge
enhance is turned on, and the bigger the value the stronger the effect.
Impacts
Larger edge enhance may lead to noise increasing.
Dynamic Range
Description
Adjusts contrast resolution of an image, compresses or expands gray display
range. The real-time dynamic range value is displayed on the image parameter
area in the upper left corner of the screen.
Operation
Adjust through the [Dyn Ra.] item in the menu
.
The adjusting range is 30-220 in increments of 5.
Effects
The more the dynamic range, the more specific the information, and the lower
the contrast with more noise.
M Soften
Description
This feature is used to process the scan lines of M images to reject noise,
making the image details to be clearer.
Operation
Adjust through the [M Soften] item in the menu;
The system provides 14 levels of M Soften adjustment, the bigger the value the
stronger the effect.
5.5 PW Doppler Mode
PW (Pulsed Wave Doppler) mode is used to provide blood flow velocity and direction
utilizing a real-time spectrum display. The horizontal axis represents time, while the vertical
axis represents Doppler frequency shift.