12-12 Probes and Biopsy
7. Check whether the probe has defects such as peeling, rifts, bumps, cracks, or liquid spill.
If such defects exist, the probe has reached the end of its service life. In this case, stop
using it and contact the Mindray service department.
NOTE:
1.
Observe the graph here carefully to immerse the transducer. Only soak parts
of the transducer below the strain relief.
2.
Repeated disinfection will eventually damage the probe, please check the
probe performance periodically.
Compatible Disinfectants
For the disinfectant information, please refer to Mindray Transducer Disinfectant
Recommendation.
Strain relief
Probe handle