15-4 EMC Guidance and Manufacturer's Declaration
TABLE 3
GUIDANCE AND MINDRAY DECLARATION—ELECTROMAGNETIC IMMUNITY
The system is intended for use in the electromagnetic environment specified below. The
customer or the user of system should assure that it is used in such an environment.
ELECTROMAGNETIC
ENVIROMENT –
GUIDANCE
Conduced RF
IEC 61000-4-6
3 Vrms
0,15 MHz – 80
MHz
6 Vrms in ISM
a
and
amateur radio
bands
a
between
0,15 MHz and 80
MHz
3 Vrms
0,15 MHz – 80
MHz
6 Vrms in ISM
a
and
amateur radio
bands
a
between
0,15 MHz and 80
MHz
Portable and mobile RF
communications equipment
should be used no closer to any
part of system, including cables,
than the recommended
separation distance calculated
from the equation applicable to
the frequency of the transmitter.
Recommended separation
distance:
d = 1.2×
80 MHz to 800
MHz
d = 2.3×
800 MHz to
2.7GHz
Where, P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey
b
,
should be less than the
compliance level in each
Radiated RF
IEC 61000-4-3