8-19 
Arrh. event  Range  Default Step  Unit 
Vtac Rate  100 to 200 
Adult, pediatric: 130 
Neonate: 160 
5 bpm 
Vtac PVCs  3 to 99  6  1  /min 
Pause Time  1.5, 2.0,2.5  2  /  s 
Vbrd PVCs  3 to 99  5  1  /min 
Vbrd Rate  15 to 60  40  5  bpm 
 
 
8.7.4 Setting the Extended Arrhythmia 
The following arrhythmia events are defined as extended arrhythmia:   
  Extreme Tachy 
  Extreme Brady 
  Vent. Brady 
  Nonsus. Vtac 
  Multif. PVC 
  Irr. Rhythm 
  Pause 
  Afib 
 
You can select [Main Menu][Maintenance >>][User Maintenance >>]enter the 
required passwordselect [Alarm Setup >>], and set [Extended Arrh.] to [Enable] or 
[Disable]. When [Extended Arrh.] is set to [Disable], the patient monitor does not analysis 
the extended arrhythmia events and corresponding alarms are not given. 
 
 
 CAUTION 
z  Set [Extended Arrh.] to [Disable] when the patient monitor is connected to the 
Central Monitoring System of version prior to 06.01.00. Failure to do so may cause 
the Central Monitoring System unable to display extended arrhythmia related 
alarms normally when extended arrhythmia occurs. 
 
 
8.7.5 Reviewing Arrhythmia Events 
Please refer to 18 Review.