EasyManuals Logo

Mindray uMEC10 User Manual

Mindray uMEC10
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #213 background imageLoading...
Page #213 background image
25-7
5. Open the relief valve to vent standard CO
2
and make sure that there is an excess gas
flow through the T-shape connector to air.
6. Check the realtime CO
2
value is within 6.0±0.3% in the [Calibrate CO2] menu.
25.6.3 Calibrating CO
2
For the CO
2
module, a calibration is needed every year or when the measured values have a
great deviation. For maintream CO
2
module, no calibration is needed. Calibration for the CO
2
module can be performed only when the module enters the full accuracy mode.
WARNING
z Connect an exhaust tube to the gas outlet connector of the monitor to remove the
calibration gases to a scavenging system.
Tools required:
 A steel gas cylinder with 6±0.05% CO
2
and balance gas N
2
 T-shape connector
 Tubing
Follow this procedure to perform a calibration:
1. Make sure that the CO
2
module has been warmed up or started up.
2. Check the airway for leakage and perform a leakage test as well to make sure the airway
has no leakage.
3. Select [Main Menu] [Maintenance >>] [User Maintenance >>] enter the
required passwordė[Module Maintenance >>] [Maintain CO
2
>>] [Calibrate
CO
2
>>].
4. In the [Calibrate CO
2
] menu, select [Zero].
5. After the zero calibration is finished successfully, connect the equipment as follows:

Table of Contents

Other manuals for Mindray uMEC10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC10 and is the answer not in the manual?

Mindray uMEC10 Specifications

General IconGeneral
Resolution800 x 600
ECG3/5-lead ECG
EtCO2Optional
BatteryRechargeable lithium-ion
Battery Life4 hours
Dimensions110 mm
AlarmAudible and visual alarms
Power SupplyAC 100-240V, 50/60Hz
AgentNo
O2No

Related product manuals