EasyManuals Logo

Mindray uMEC10 User Manual

Mindray uMEC10
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #277 background imageLoading...
Page #277 background image
E-3
E.4 Protective Earth Resistance
1. Plug the probes of the analyzer into the device’s protective earth terminal and protective
earth terminal of the AC power cord.
2. Test the earth resistance with a current of 25 A.
3. Verify the resistance is less than limits.
LIMITS
For all countries, R = 0.2  Maximum
E.5 Earth Leakage Test
Run an Earth Leakage test on the device being tested before performing any other leakage
tests.
The following outlet conditions apply when performing the Earth Leakage test:
 normal polarity (Normal Condition)
 reverse polarity (Normal Condition)
 normal polarity with open neutral (Single Fault Condition)
 reverse polarity with open neutral (Single Fault Condition)
LIMITS
For UL60601-1,
 300 A in Normal Condition
 1000 A in Single Fault Condition
For IEC60601-1,
 500 A in Normal Condition
 1000 A in Single Fault Condition

Table of Contents

Other manuals for Mindray uMEC10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC10 and is the answer not in the manual?

Mindray uMEC10 Specifications

General IconGeneral
Resolution800 x 600
ECG3/5-lead ECG
EtCO2Optional
BatteryRechargeable lithium-ion
Battery Life4 hours
Dimensions110 mm
AlarmAudible and visual alarms
Power SupplyAC 100-240V, 50/60Hz
AgentNo
O2No

Related product manuals