7-5
7.4 Alarm Tone Configuration
7.4.1 Setting the Minimum Alarm Volume
1. Select [Main Menu][Maintenance >>][User Maintenance >>]enter the
required password.
2. Select [Alarm Setup >>] to enter the [Alarm Setup] menu.
3. Select [Minimum Alarm Volume] and toggle between 0 and 10.
The minimum alarm volume refers to the minimum value you can set for the alarm volume,
which is not affected by user or factory default configurations. The setting of minimum
alarm volume remains unchanged when the patient monitor shuts down and restarts.
7.4.2 Changing the Alarm Volume
1. Select the [Volume Setup] QuickKey or the [Alarm Setup] QuickKeyÄ—[Others], or
[Main Menu][Alarm Setup >>]ė[Others].
2. Select the appropriate volume from [Alm Volume]: X-10, in which X is the minimum
volume, depending on the set minimum alarm volume, and 10 the maximum volume.
3. Select [High Alarm Volume] to set the volume of the high priority alarm as [Alm
Volume+0], [Alm Volume+1] or [Alm Volume+2].
4. Select [Reminder Vol] to set the volume of the reminder tone as [High], [Med] or
[Low].
When alarm volume is set to 0, the alarm sound is turned off and a
symbol appears on
the screen.
7.4.3 Setting the Interval between Alarm Sounds
You cannot change the interval between alarm tones if you choose mode 1 or 2 as your
desired alarm tone pattern. For these two patterns, the interval between alarm tones
identifies the alarm levels as follows:
 Mode 1:
 Interval between high level alarm tones: continuously.
 Interval between medium level alarm tones: 5 s.
 Interval between low level alarm tones: 20 s.

Table of Contents

Other manuals for Mindray uMEC10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC10 and is the answer not in the manual?

Mindray uMEC10 Specifications

General IconGeneral
Resolution800 x 600
ECG3/5-lead ECG
EtCO2Optional
BatteryRechargeable lithium-ion
Battery Life4 hours
Dimensions110 mm
AlarmAudible and visual alarms
Power SupplyAC 100-240V, 50/60Hz
AgentNo
O2No

Related product manuals