21.6 Setting Up Reports
21.6.1 Setting Up ECG Reports
You can print out ECG reports only under full-screen, half-screen or 12-lead monitoring
screen. To set up ECG reports, select [Main Menu]Ä—[Print Setup >>]Ä—[ECG
Reports >>].
 [Amplitude]: set the amplitude of the ECG waveforms.
 [Sweep]: set the wave print speed to 25 mm/s or 50 mm/s.
 [Auto Interval]: If [Auto Interval] is set to [On], the system will automatically adjust
the space between waveforms to avoid overlapping.
 [12-Lead Format]: If you select [12
1], 12 waveforms will be printed on a paper from
top to bottom. If you select [6
2], 12 waveforms will be printed from left to right with
6 waveforms on each half part and a rhythm waveform will be printed at the bottommost.
If you select [3
4+1], 12 waveforms will be printed from left to right with 3
waveforms on each of the 4 columns and a rhythm waveform will be printed at the
bottommost.
21.6.2 Setting Up Tabular Trends Reports
To set up tabular trends reports, select [Main Menu]Ä—[Print Setup >>]Ä—[Tabular Trends
Reports >>].
 Date: Select the desired date on which a tabular trends report is recorded for printing.
 Start time: You can set a time period whose trend data will be printed out by setting
[From] and [Back]. For example, if you set [From] as 2007-4-2 10: 00: 00 and [Back]
as [2 h], the outputted data will be from 2007-4-2 08: 00: 00 to 2007-4-2 10: 00: 00. In
addition, the [Back] can be set to either:
 [Auto]: If [Report Layout] is set to [Time Oriented], the report will be printed by
time. If [Report Layout] is set to [Parameter Oriented], the report will be printed
by parameters.
 [All]: If you select [All], all trend data will be printed out. In this case, it is no need
to set [From].
 [Spacing]: choose the resolution of the tabular trends printed on the report.
 [NIBP]: prints a tabular trends report when NIBP measurements were acquired.
Each NIBP measurement will generate a column in the tabular trends review
window. The values for the other measurements are added to the column to provide
a complete dataset for the NIBP measurement time.
 [Auto]: prints a tabular trends report during the period selected and at the interval
set in the [Interval] option. The [Interval] option is located at the bottom the
[Tabular Trends] tab of the [Review] window.