EasyManua.ls Logo

Mindray uMEC10 - Power Supply Specifications

Mindray uMEC10
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A-2
NOTE
z The environmental specifications of unspecified parameters are the same as those
of the main unit.
A.2 Power Supply Specifications
AC power
Line voltage 100 to 240 VAC (±10ˁ)
Current 1.5 A
Frequency 50/60 Hz (±3Hz)
Battery (standard configuration)
Battery Type Chargeable Lithium-Ion, 11.1DVC, 2.5 Ah
Run time
uMEC15:  2.5 h
uMEC10/uMEC12:  4 h
when powered by a new fully-charged battery (25ć, display
brightness set to default value, SpO
2
cable connected, ECG
cable and Temp cable not connected, auto NIBP measurements
at an interval of 15 minutes)
Charge time
When the monitor is powered off,
less than 3.5 hours to 90%
less than 4 hours to 100%
When the monitor is in use,
less than 5.5 hours to 90%
less than 6 hours to 100%
Shutdown delay at least 5 min (after a low battery alarm first occurs)
Battery (optional configuration)
Battery Type Chargeable Lithium-Ion, 11.1DVC, 5 Ah
Run time
uMEC15:  6 h
uMEC10/uMEC12:  8 h
when powered by a new fully-charged battery (25ć, display
brightness set to default value, SpO
2
cable connected, ECG
cable and Temp cable not connected, auto NIBP measurements
at an interval of 15 minutes)
Charge time
When the monitor is powered off,
Less than 7 hours to 90%
Less than 8 hours to 100%