7-12
7.5.3 Setting Alarm Delay Time
You can set the alarm delay time for over-limit alarms of continuously measured parameters.
If the alarm-triggered condition disappears within the delay time, the patient monitor will
not give the alarm. You can set the [Alarm Delay] in the [Others] window of [Alarm
Setup] menu.
To set the alarm delay time,
1. Select [Main Menu] [Maintenance >>] [User Maintenance >>]. Enter the
required password and then select [Ok].
2. Select [Alarm Setup >>][Alarm Delay].
Alarm delay is not applied to the following physiological alarms:
 Apnea
 ST alarms
 Arrhythmia alarms
 ECG Weak Signal
 Resp Artifact
 No Pulse
 Measurements of noncontinuous parameters over alarm limits
 HR over alarm limits
You can set [Apnea Delay] and [ST Alarm Delay] separately.
7.5.4 Setting SpO
2
Technical Alarm Delay
You can set [Tech. Alarm Delay] in the [Others] tab of the [Alarm Setup] menu. The
options are [Off], [5s], [10s] and [15s]. The delay is effective to the following technical
alarms: SpO
2
Sensor Off, SpO
2
Too Much Light, SpO
2
Low Signal and SpO
2
Interference.

Table of Contents

Other manuals for Mindray uMEC10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC10 and is the answer not in the manual?

Mindray uMEC10 Specifications

General IconGeneral
Resolution800 x 600
ECG3/5-lead ECG
EtCO2Optional
BatteryRechargeable lithium-ion
Battery Life4 hours
Dimensions110 mm
AlarmAudible and visual alarms
Power SupplyAC 100-240V, 50/60Hz
AgentNo
O2No

Related product manuals