EasyManuals Logo

Mindray uMEC10 User Manual

Mindray uMEC10
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #83 background imageLoading...
Page #83 background image
7-11
Module Parameter
Low alarm limit High alarm limit
Auto alarm limits
range
Adult/
pediatric
Neonate
Adult/
pediatric
Neonate
CO
2
EtCO
2
0 to 32mmHg:
remains the
same
0 to
32mmHg:
remains the
same
0 to 32mmHg:
remains the
same
0 to
32mmHg:
remains the
same
Same as the
measurement range
32 to
35mmHg:
29mmHg
32 to
35mmHg:
29mmHg
32 to
35mmHg:
41mmHg
32 to
35mmHg:
41mmHg
35 to
45mmHg:
(etCO
2
-6)
mmHg
35 to
45mmHg:
(etCO
2
-6)
mmHg
35 to
45mmHg:
(etCO
2
+6)
mmHg
35 to
45mmHg:
(etCO
2
+6)
mmHg
45 to
48mmHg: 39
mmHg
45 to
48mmHg:
39 mmHg
45 to
48mmHg: 51
mmHg
45 to
48mmHg:
51 mmHg
>48mmHg:
remains the
same
>48mmHg:
remains the
same
>48mmHg:
remains the
same
>48mmHg:
remains the
same
FiCO
2
N/A N/A
Same as the
default alarm
limit
Same as the
default
alarm limit
Same as the
measurement range
awRR
awRR × 0.5 or
6 rpm
(whichever is
greater)
(awRR – 10)
or 30 rpm
(whichever
is greater)
awRR × 1.5 or
30 rpm
(whichever is
smaller)
(awRR+25)
or 85 rpm
(whichever
is smaller)
Adult/pediatric: 6 to
55
Neonate: 10 to 90
C.O. BT
Adult:
(BT – 1)ć
N/A
Adult:
(BT – 1)ć
N/A
Same as the
measurement range

Table of Contents

Other manuals for Mindray uMEC10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC10 and is the answer not in the manual?

Mindray uMEC10 Specifications

General IconGeneral
Resolution800 x 600
ECG3/5-lead ECG
EtCO2Optional
BatteryRechargeable lithium-ion
Battery Life4 hours
Dimensions110 mm
AlarmAudible and visual alarms
Power SupplyAC 100-240V, 50/60Hz
AgentNo
O2No

Related product manuals